Skip to content

Functional Overview

This chapter provides the overview of the features of the MFIA Instrument. The first section contains the graphical overview as well as the hardware and software feature list. The next sections detail the front panel and the back panel of the measurement instrument. The following section describes the MFITF Impedance Test Fixture. The last section provides product selection and ordering support.

Features

Figure 1: MFIA Instrument Overview

Every MF Instrument (MFLI, MFIA) described on Figure 1 consists of several internal units (light blue color) surrounded by several interface units (dark blue color) and the front panel on the left-hand side and the back panel on the right-hand side. The orange blocks are optional units that can be either ordered at the beginning or upgraded later in the field. The arrows between the panels and the interface units indicate the physical connections and the data direction flow. Only a small subset of internal connections is depicted.

The MFIA 5 MHz Precision LCR Meter and Impedance Analyzer provides fast and accurate measurements over a wide impedance range. Unlike other impedance analyzers the MFIA measures the voltage and the current signals directly without the need of a feedback loop, e.g. balanced bridge configuration. As a result the MFIA is capable of measuring at frequencies as low as 1 Hz and to provide reliable impedance measurements up to 10 GΩ.

The device-under-test is connected to the low-noise inputs and outputs LCUR, HCUR, LPOT, HPOT via the MFITF Impedance Test fixture. If a different test fixture is used, the Compensation Advisor makes it easy to obtain accurate measurement data. The measurement signals are amplified to an appropriate range and digitized at very high speed. The resulting samples are fed into the digital signal processor with dual-phase demodulators for independent current and voltage measurement. The measurement signals are further processed on the embedded processor that provides the LabOne Web Server. Any device running a browser (PC, tablet or smart phone) can connect to the Web Server and no software installation is required. Both Ethernet and USB interfaces are supported. The measurement data are also available as analog signal on the auxiliary outputs on the front panel of the MF Instrument. When the MF-MD option is installed, the drive output HCUR is capable of generating a multi-frequency output signal which enables impedance measurements at a second independent frequency.

The MFIA Impedance Analyzer provides full lock-in functionality equivalent to the MFLI Lock-in Amplifier. Several trigger and reference connections provide the associated functionality for purposes such as triggering demodulation, triggering oscilloscope data acquisition, or to generate external reference clocks or triggering signals to other equipment.

Impedance Analyzer features

  • 2-terminal and 4-terminal measurement modes
  • Multiple DUT representations (replacement circuits)
  • Confidence Indicator
  • Compensation Advisor
  • Auto range mode
  • Auto bandwidth mode
  • Auto test signal level
  • Dual-frequency mode (requires MF-MD option, simultaneous impedance measurement at two frequencies)

Medium Frequency Voltage Input LPOT, HPOT

  • 1 low-noise MF voltage input, single-ended or differential, 5 MHz bandwidth
  • Variable input range
  • Switchable input impedance
  • Selectable AC/DC coupling
  • Selectable instrument ground or floating

Medium Frequency Current Input LCUR

  • 1 low-noise MF current input, single-ended, 5 MHz bandwidth
  • Variable input range

Medium Frequency Signal Output HCUR

  • Low-distortion MF outputs, single-ended, differential, 5 MHz bandwidth
  • Variable output range
  • Digital and analog offset

Auxiliary Input, Outputs and Triggers

  • 4 auxiliary outputs, user defined signals
  • 2 auxiliary inputs, general purpose
  • 2 input and 2 output trigger signals

High-speed Connectivity

  • USB 2.0 Device high-speed 480 Mbit/s interface
  • Dual USB 2.0 Host high-speed interface
  • LAN 1 Gbit/s controller interface
  • DIO: 32-bit digital input-output port
  • Clock input connector (10 MHz)
  • Clock output connector (10 MHz)

LabOne Toolset

  • Sweeper
  • Scope
  • Numeric
  • Spectrum
  • Plotter
  • Data Acquisition

Software Connectivity

  • Data server with multi-client support
  • API for C, LabVIEW, MATLAB, Python based instrument programming

Front Panel Tour

The front panel BNC connectors and control LEDs are arranged as shown in Figure 2 and listed in Table 1.

Note

Some BNC connectors on the front panel have secondary labels below the connector. These are relevant when operating the MFIA as a lock-in amplifier and are listed in brackets in Table 1.

Figure 2: MF Instrument front panel

Table 1: MF Instrument front panel description
Position Label / Name Description
A Signal Input LCUR (I) single-ended current input
B Current Input Signal Over this red LED indicates that the current input signal saturates the A/D converter and therefore the current input range must be increased or the signal must be attenuated
C Signal Input LPOT (–V Diff) voltage input
single-ended mode: internally shorted to ground
differential mode: negative voltage input
D Signal Input HPOT (+V) voltage input
in single-ended mode: single-ended voltage input
differential mode: positive voltage input
E Voltage Input Signal Over this red LED indicates that the voltage input signal saturates the A/D converter and therefore the voltage input range must be increased or the signal must be attenuated
F Signal Output HCUR(+V) Drive voltage output
single-ended mode: single-ended voltage output
differential mode: positive voltage output
G Signal Output –V Diff voltage output
single-ended mode: internally shorted to ground
differential mode: negative voltage output
H Signal Output ON this blue LED indicates that the signal output is actively driven by the instrument
I Aux Input 2 Ref auxiliary input 2, can be used as external reference input supporting the full bandwidth of the device
J Aux Output 2 auxiliary output 2, this connector provides a user defined signal, often used to output demodulated samples (X,Y) or (R,THETA)
K Aux Output 4 auxiliary output 4, this connector provides a user defined signal, often used to output demodulated samples (X,Y) or (R,THETA)
L Power this LED indicates that the instrument is powered
color blue: the device has an active connection to the LabOne data server and is ready for operation
color green blinking: the firmware is ready, waiting for LabOne data server connection. This process takes around 20 seconds.
color red: the device is not initialized respectively is performing the internal auto calibration process.
color purple blinking: a firmware update is in progress
color purple: the boot process failed
M Aux Input 1 Ref auxiliary input 1, can be used as external reference input supporting the full bandwidth of the instrument; the value of auxiliary input 1 can be added as offset to the signal output
N Aux Output 1 auxiliary output 1, this connector provides a user-defined signal, often used to output demodulated samples (X,Y) or (R,Θ)
O Aux Output 3 auxiliary output 3, this connector provides a user-defined signal, often used to output demodulated samples (X,Y) or (R,Θ)

Please refer to the troubleshooting section for further information regarding the meaning of the Power LED colors.

Back Panel Tour

The back panel is the main interface for power, control, service and connectivity to other ZI instruments. Please refer to Figure 3 and Table 2 for the detailed description of the items.

Figure 3: MF Instrument back panel

Table 2: MF Instrument back panel description
Position Label / Name Description
A DC In DC external 12 V power supply
B DIO 32-bit digital input/output connector
C USB 2.0 Host universal serial bus host connector
D USB 2.0 Device universal serial bus device connector to computer
E LAN 1GbE 1 Gbit Ethernet LAN connector
F Trigger In 2 digital TTL trigger input 2
G Trigger Out 2 digital TTL trigger output 2
H Clk 10 MHz In clock input (10 MHz) to be used for synchronization from external instruments
I Clk 10 MHz Out clock output (10 MHz) to be used for synchronization of external instruments
J Earth ground 4 mm banana jack connector for earth ground, electrically connected to the chassis and the earth pin of the power inlet
K USB 2.0 Host universal serial bus host connector
L Power inlet power inlet with ON/OFF switch
M Trigger In 1 digital TTL trigger input 1
N Trigger Out 1 digital TTL trigger output 1
O Cooling outlet ventilator (important: keep clear from obstruction)

MFITF Impedance Test Fixture

The MFIA Impedance Analyzer is sold in combination with the MFITF Impedance Test Fixture. The test fixture shown in Figure 4 is to be directly connected to the LCUR, LPOT, HPOT, HCUR BNC connectors on the MFIA front panel. The spacing of the BNC connectors is compatible with established standards for LCR meter and impedance analyzer test fixtures. The MFITF comes with a set of low-capacitive and low-loss DUT carriers for surface-mounted devices (SMD) and for through-hole components. Using the MFIA together with the MFITF and these carriers ensures the best possible accuracy.

Figure 4: MFITF Impedance Test Fixture top view

The MFITF front panel as shown in Figure 5 integrates an 8-pin connector (Sullins PPPC081LFBN-RC, Digikey part number S7041-ND) providing four signal contacts and four guard contacts for customized DUT holders.

Figure 5: MFITF Impedance Test Fixture front view

The PCB-based carriers delivered with the MFITF are described in the following table.

Table 3: DUT carriers for MFITF Impedance Test Fixture
Label Image Description Number of units
Open Open circuit for compensation and testing 1
Short Short-circuit for compensation and testing 1
1k 0.05% 1 kΩ load with 0.05% accuracy, 10 ppm/°C temperature coefficient, for testing and compensation 1
2 Pt Carrier for SMDs with 2-terminal connection 3
4 Pt Carrier for SMDs with 4-terminal connection 4
Through hole Carrier for through-hole components 2

Ordering Guide

Table 4 provides an overview of the available MF products. Upgrade options can be purchased at any time without need to send the Instrument to Zurich Instruments.

Table 4: MF Instrument product codes for ordering
Product code Product name Description Upgrade in the field possible
MFIA 500 kHz MFIA 500 kHz Lock-in Amplifier base product -
MFIA 5 MHz MFIA 5 MHz Lock-in Amplifier bundle -
MF-MD MF-MD Multi-demodulator option yes
MF-F5M MF-F5M Frequency Extension option yes
MF-DIG MF-DIG Digitizer option yes
MF-IA MF-IA Impedance Analyzer option (includes MFITF) yes
MFITF Impedance Test Fixture accessory (included with the MFIA and MF-IA) -

The MF product line consists of the MFLI Lock-in Amplifier series and the MFIA Impedance Analyzer series. The combination MFLI 5 MHz + MF-IA is equivalent to the MFIA 5 MHz product, and the combination MFLI 500 kHz + MF-IA is equivalent to the MFIA 500 kHz product. The following table gives an overview of the features for the most important product configurations.

Table 5: Product selector
Feature MFIA MFIA +MF-MD MFIA +MF-F5M MFIA +MF-MD +MF-F5M
Internal reference mode yes yes yes yes
External reference mode yes yes yes yes
Auto reference mode yes yes yes yes
Impedance mode (independent measurement of voltage and current signal inputs) - yes - yes
Signal generators 1 1 1 1
Superposed output sinusoids per generator 1 up to 4 1 up to 4
Quad-harmonic mode - yes - yes
Multi-frequency mode - yes - yes
Arbitrary frequency mode - yes - yes
Number of demodulators 1 4 1 4
Simultaneous frequencies 1 4 1 4
Simultaneous harmonics 1 4 1 4
External references 1 2 1 2
Dynamic reserve 120 dB 120 dB 120 dB 120 dB
Frequency range 500 kHz 500 kHz 5 MHz 5 MHz
USB 2.0 480 Mbit/s yes yes yes yes
Ethernet 1 GbE yes yes yes yes

Rack Mount

Parts for mounting the MFIA in a 19" rack are available from Zurich Instruments upon request under support@zhinst.com. Rack mount of a single MFIA as well as two MFIA instruments side-by-side is possible and explained in the following instructions.

Instructions for a Single Instrument

Table 6: Parts for rack mount of single instrument
Description Quantity Unit
19" Corner bracket Magic 2U RAL 9003 1 PCE
19" Adapter Plate 2U RAL 9003 1 PCE
Pan head screw M6 x 16 ISO 14583 A2 T30 4 PCE
DIN 125 A 6.4 st. zinc plated washer 4 PCE
Cage nut KM 6 4 PCE
Screw ISO 7380, M4x12, Torx, zinc plated 2 PCE
Washer DIN 9021 Ø 4.3 zinc plated 2 PCE
  1. Remove 4x screws M3x7

  2. Remove 2x trim piece

  3. Attach 2x 19" Corner bracket Magic 2U RAL 9003 using the screws which were removed in step 1

  4. Attach 19"Adapter plate 2U RAL 9003 using 2x screw ISO 7380 M4x12 Torx and washer DIN 9021 Ø 4.3

  5. Attach 4x washer DIN 125 A 6.4 and 4x screw ISO 14583 M6x16 Torx for rack mount

  6. Attach 4x cage nuts KM6 to cabinet

  7. Attach instrument to cabinet

Instructions for Two Instruments

Table 7: Parts for rack mount of two instrument
Description Quantity Unit
Connection Strap Magic 2U outside 1 PCE
Connection Strap Magic 2U inside 1 PCE
19" Corner bracket Magic 2U RAL 9003 2 PCE
Connection Strap Magic 2U backside 1 PCE
Screw sim. DIN 965 M2.5x10 A2 Torx lock 2 PCE
Screw sim. DIN 7985 M3x12 Torx A2 w. lock-tight 2 PCE
Pan head screw M6 x 16 ISO 14583 A2 T30 4 PCE
DIN 125 A 6.4 st. zinc plated washer 4 PCE
Cage nut KM 6 4 PCE

Assembly Instructions: Left Chassis

  1. Remove 2x screws M3x7 on the left side and remove left trim piece

  2. Remove 2x screws M3x7 on the right side

  3. Attach 19" Corner bracket Magic 2U RAL 9003 to the left side with 2x screw M3x7 removed in step 1

  4. Attach Connection Strap Magic 2U outside to the right side with 2x DIN 7985 screw M3x12 Torx

Assembly Instructions: Right Chassis

  1. Remove 2x screws M3x7 on the right side and remove right trim piece

  2. Remove 2x screws M3x7 left side

  3. Attach 19" Corner bracket Magic 2U RAL 9003 to the right side with 2x screw M3x7 removed in step 1

  4. Attach Connection Strap Magic 2U inside to left the side with screw Cylinder head screw M3x12 Torx

Chassis Assembly

  1. Attach both chassis together and fix with 2 screws DIN 965 M2.5x10 Torx through the connection strap

  2. Remove 4x screws and 2x inner plastic covers

  3. Attach the rear connection strap, using screws removed in step 2

  4. Attach Cage nuts KM6 to cabinet