Ports Tab

The Ports tab provides control and displays the status of the 56 ZSync ports. It is available on all QHub instruments.

Features

  • Display the state of the complete system at first glance
  • Display port synchronization of all 56 ZSync ports
  • Control of triggers that are sent to the connected instruments

Description

Table 1: App icon and short description
Control/Tool Option/Range Description
Ports Control and status display of the ZSync ports.

The Ports tab (see LabOne UI: Ports tab) is divided into two sections: The ZSync ports and two sub-tabs for Control and Trigger.

Figure 1: LabOne UI: Ports tab

The QHub is always used in conjunction with other Zurich Instruments devices in a larger system, e.g. a Quantum Computing Control System (QCCS). The purpose of the Ports tab is to allow the user to understand the current state of the complete system at first glance. The main elements are the 56 ports, organized in four sub-tabs, where the user can see information about the status and health of each connection and can reset the connection. Furthermore the user can start the sending of the triggers and control the repetitions and the holdoff of the triggers.

Functional Elements

Table 2: Ports tab
Control/Tool Option/Range Description
Connection Status off/blue/yellow/red Indicates the availability of the instrument connected to the port. Off: no Instrument detected. Yellow: connection to an instrument is in progress. Blue: connection to an instrument is ready or data is being sent to / received from an instrument. Red: an error has occurred on the connection to an instrument.
Serial The device ID of the instrument connected to this port.
Device Type The device type of the instrument connected to this port.
Run/Stop Starts sending triggers to all connected instruments over ZSync ports.
Repetitions Sets the number of triggers sent over ZSync ports.
Holdoff time in seconds Sets the time between repeated triggers sent over ZSync ports.
Progress 0% to 100% The percentage of repeated triggers sent over ZSync ports.
Synchronization Enable synchronization. Trigger generation will only start once all workers have reported a ready status. Synchronization checks will be repeated with the same trigger generation settings (holdoff and repetitions) until synchronization is disabled.
Enable ON / OFF Enable Trigger Out connector.
Source Port Select the ZSync port associated with the Trigger Out source.