General Specifications
Table 1: General and storage
Parameter |
Min |
Typ |
Max |
storage temperature |
–25 °C |
- |
65 °C |
storage relative humidity (non-condensing) |
- |
- |
95% |
operating temperature |
5 °C |
- |
40 °C |
operating relative humidity (non-condensing) |
- |
- |
90% |
specification temperature |
18 °C |
- |
28 °C |
power consumption |
- |
- |
120 W |
operating environment |
IEC61010, indoor location, installation category II, pollution degree 2 |
|
|
operating altitude |
up to 2000 meters |
|
|
power inlet fuses |
250 V, 2 A, fast acting, 5 x 20 mm |
|
|
power supply AC line |
100-240 V (±10%), 50/60 Hz |
|
|
dimensions (width x depth x height) |
44.0 × 38.0 × 4.4 cm (no handle), 17.3 × 15.0 × 1.7 inch, 19 inch rack compatible |
|
|
weight |
6 kg (14 lb) |
|
|
recommended calibration interval |
2 years |
|
|
Table 2: Maximum ratings
Parameter |
Details |
Min |
Typ |
Max |
Damage threshold Pump Out |
voltage level |
-15 V |
- |
+15 V |
Damage threshold Pump In |
voltage level |
-15 V |
- |
+15 V |
|
RF power |
- |
- |
+20 dBm |
Damage threshold Signal Out |
voltage level |
-15 V |
- |
+15 V |
Damage threshold Signal In |
voltage level |
-15 V |
- |
+15 V |
|
RF power |
- |
- |
+20 dBm |
Damage threshold Trig Out |
voltage level |
-0.5 V |
- |
+3.8 V |
Damage threshold Trig In |
voltage level (50 Ω input impedance) |
-0.5 V |
- |
+5.5 V |
Damage threshold DIO Out |
voltage level |
-0.5 V |
- |
+3.8 V |
Damage threshold DIO In |
voltage level |
-0.5 V |
- |
+6.5 V |
Damage threshold External Clk Out |
voltage level |
-3 V |
- |
+3 V |
Damage threshold External Clk In |
voltage level |
-15 V |
- |
+15 V |
|
RF power |
- |
- |
+15 dBm |
Table 3: Host computer requirements
Parameter |
Description |
supported Windows operating systems |
Windows 10, 11 on x86-64 |
supported macOS operating systems |
macOS 10.11+ on x86-64 and ARMv8 |
supported Linux distributions |
GNU/Linux (Ubuntu 14.04+, CentOS 7+, Debian 8+) on x86-64 and ARMv8 |
supported processors |
x86-64 (Intel, AMD), ARMv8 (e.g., Raspberry Pi 4 and newer, Apple M-series) |
Analog Interface Specifications
Table 4: Pump Tone Generation (Parametric Pump Channels)
Parameter |
Details |
|
Min |
Typ |
Max |
Output impedance |
- |
|
- |
50 Ω |
- |
Output coupling |
- |
|
AC |
|
|
Synthesizer frequency range |
Pump Tone |
|
- |
4...12 GHz |
2.5...12 GHz |
|
Probe Tone |
|
- |
4...8.5 GHz |
2.5...10 GHz |
Synthesizer frequency step |
- |
|
1 kHz |
- |
- |
Nominal output level range |
Pump Tone |
HIGH |
- |
0...20 dBm |
- |
|
|
LOW |
- |
-20...0 dBm |
- |
|
Probe Tone |
at 6 GHz |
- |
-35...-50 dBc |
- |
Maximum output level |
Pump Tone, ALC enabled and low-pass filter disabled |
4…8 GHz |
- |
20 dBm |
- |
|
|
8…9 GHz |
- |
18 dBm |
- |
|
|
9…10 GHz |
- |
14 dBm |
- |
|
|
10…11 GHz |
- |
12 dBm |
- |
|
|
11…12 GHz |
- |
8 dBm |
- |
Output level error |
Pump Tone |
|
- |
1.5 dB |
- |
|
Probe Tone |
|
- |
3 dB |
- |
Output level step |
Pump Tone |
ALC enabled |
- |
0.01 dB |
- |
|
Probe Tone |
|
- |
0.5 dB |
- |
Maximum output level drift with temperature |
Pump Tone |
ALC enabled |
- |
0.02 dB / °C |
- |
Maximum output level drift over time |
Pump Tone |
ALC enabled, at 25 °C |
- |
0.02 dB / 24h |
- |
Output phase noise |
10 GHz |
offset 1 kHz |
- |
-95 dBc/Hz |
- |
|
|
offset 10 kHz |
- |
-105 dBc/Hz |
- |
|
|
offset 1 MHz |
- |
-115 dBc/Hz |
- |
|
|
offset 10 MHz |
- |
-145 dBc/Hz |
- |
Table 5: Readout Signal Conditioning (Parametric Pump Channels)
Parameter |
Details |
Min |
Typ |
Max |
Signal In impedance |
- |
- |
50 Ω |
- |
Signal In coupling |
- |
AC |
|
|
Signal Out impedance |
- |
- |
50 Ω |
- |
Signal Out coupling |
- |
AC |
|
|
Pump In impedance |
- |
- |
50 Ω |
- |
Pump In coupling |
- |
AC |
|
|
Signal In to Signal Out frequency range |
- |
- |
5...9 GHz |
- |
Signal In to Signal Out gain |
- |
- |
10 dB |
- |
Signal In to Signal Out noise figure |
8 GHz |
- |
5 dB |
- |
Signal In maximum level |
8 GHz |
- |
-7 dBm |
- |
Pump tone cancellation frequency range |
- |
- |
5...10 GHz |
- |
Cancellation tone attenuation span |
- |
- |
50 dB |
- |
Cancellation tone phase shift span |
- |
360 deg |
- |
- |
Table 6: Other Inputs and Outputs
Parameter |
Details |
Min |
Typ |
Max |
Reference clock input |
- |
SMA on back panel |
|
|
Reference clock input impedance |
- |
- |
50 Ω |
- |
Reference clock input coupling |
- |
AC |
|
|
Reference clock input frequency |
- |
10 / 100 MHz |
|
|
Reference clock input level range |
- |
- |
-10...+14 dBm |
- |
Reference clock output |
- |
SMA on back panel |
|
|
Reference clock output impedance |
- |
- |
50 Ω |
- |
Reference clock output coupling |
- |
AC |
|
|
Reference clock output frequency |
- |
10/100 MHz |
|
|
Reference clock output level |
10MHz |
- |
+8 dBm |
- |
|
100 MHz |
- |
+7 dBm |
- |