Specifications
Important
Unless otherwise stated, all specifications apply after 30 minutes of
instrument warm-up.
Important
Important changes in the specification parameters are explicitly
mentioned in the revision history of this document.
QCCS system specifications (systems based on SHF series and HDAWG)
Parameter |
min |
typ |
max |
Supported peripheral devices |
SHFQC, HDAWG, SHFSG, SHFQA |
|
|
Supported system architecture |
Star architecture |
|
|
Output channel skew |
- |
- |
1.6 ns |
Output channel skew repeatability over restarts |
- |
- |
200 ps |
Real-time communication latency |
- |
- |
90 ns |
Sample rate |
- |
2.0 GSa/s |
- |
QCCS system specifications (systems based on UHFQA and HDAWG)
Parameter |
min |
typ |
max |
Supported peripheral devices |
HDAWG, UHFQA |
|
|
Supported system architecture |
Star architecture |
|
|
HDAWG Output channel skew |
- |
- |
1.6 ns |
HDAWG Output channel skew repeatability over restarts |
- |
- |
200 ps |
Real-time communication latency |
- |
- |
330 ns from UHFQA to PQSC 100 ns from PQSC to HDAWG |
Sample rate |
- |
HDAWG 2.4 GSa/s UHFQA 1.8 GSa/s |
- |
This configuration requires one HDAWG per UHFQA as bridge over the DIO port of the HDAWG.
Synchronization and real-time data exchange interface
Parameter |
Value |
Interface type |
ZSync proprietary |
Number of ports |
18 |
Cable length |
3m |
Analog Interface Specifications
Table 2: Trigger and other outputs
Parameter |
Details |
min |
typ |
max |
trigger outputs |
- |
2 SMA on back panel |
|
|
trigger output impedance |
- |
50 Ω |
|
|
trigger output voltage range |
50 Ω impedance |
0 V |
- |
3.3 V |
reference clock output |
- |
SMA on back panel |
|
|
reference clock output amplitude |
100 MHz into 50 Ω |
- |
+4dBm |
- |
reference clock output frequency |
- |
Auto-detect 10 MHz / 100 MHz |
|
|
General Specifications
Table 3: General and storage
Parameter |
min |
typ |
max |
storage temperature |
–25 °C |
- |
65 °C |
storage relative humidity (non-condensing) |
- |
- |
95% |
operating temperature |
5 °C |
- |
40 °C |
operating relative humidity (non-condensing) |
- |
- |
90% |
specification temperature |
18 °C |
- |
28 °C |
power consumption |
- |
- |
100 W |
operating environment |
IEC61010, indoor location, installation category II, pollution degree 2 |
|
|
operating altitude |
up to 2000 meters |
|
|
power supply AC line |
100-240 V (±10%), 50/60 Hz |
|
|
dimensions with handles and feet |
45.0 × 34.5 × 10.0 cm, 17.7 × 13.6 × 3.9 inch, 19 inch rack compatible |
|
|
weight |
6.0 kg |
|
|
recommended calibration interval |
2 years |
|
|
Table 4: Maximum ratings
Parameter |
min |
typ |
max |
damage threshold Trigger Out 1 and 2 |
–0.7 V |
– |
+4 V |
damage threshold Trigger In 1 and 2 |
–0.7 V |
– |
+4 V |
damage threshold Reference Clock Out (DC) |
-4 V |
– |
+4 V |
damage threshold Reference Clock In (AC, with DC offset 0 V) |
– |
– |
+13.5 dBm |
damage threshold Reference Clock In (DC) |
-4 V |
– |
+4 V |
damage threshold DIO In / Out port |
-0.7 V |
– |
+4 V |
Table 5: Host computer requirements
Parameter |
Description |
supported Windows operating systems |
Windows 10, 11 on x86-64 |
supported macOS operating systems |
macOS 10.11+ on x86-64 and ARMv8 |
supported Linux distributions |
GNU/Linux (Ubuntu 14.04+, CentOS 7+, Debian 8+) on x86-64 and ARMv8 |
supported processors |
x86-64 (Intel, AMD), ARMv8 (e.g., Raspberry Pi 4 and newer, Apple M-series) |
Table 6: Digital Interface Specifications
Parameter |
Description |
host computer connection |
1GbE, LAN / Ethernet |
|
USB 3.0 |
Debug interface |
JTAG (over USB) |